Human DNA repair protein XRCC2 (XRCC2) ELISA Kit
Référence AE10989HU-48
Conditionnement : 48Tests
Marque : Abebio
Reagent Preparation

Results demonstration

Assay Procedure Summary

Certificate

Product Details
Species Reactivity | Human (Homo sapiens) |
UniProt | O43543 |
Abbreviation | XRCC2 |
Alternative Names | DKFZp781P0919; DNA repair protein XRCC2|RAD51-like|X-ray repair cross complementing protein 2|X-ray repair; complementing defective; repair in Chinese hamster |
Range | 0.312-20 ng/mL |
Sensitivity | 0.114 ng/mL |
Sample Type | Serum, Plasma, Other biological fluids |
Detection Method | Sandwich |
Analysis Method | Quantitive |
Assay Duration | 1-4.5h |
Sample Volume | 1-200 μL |
Detection Wavelengt | 450 nm |
Test principle
This assay employs a two-site sandwich ELISA to quantitate XRCC2 in samples. An antibody specific for XRCC2 has been pre-coated onto a microplate. Standards and samples are pipetted into the wells and any XRCC2 present is bound by the immobilized antibody. After removing any unbound substances, a biotin-conjugated antibody specific for XRCC2 is added to the wells. After washing, Streptavidin conjugated Horseradish Peroxidase (HRP) is added to the wells. Following a wash to remove any unbound avidin-enzyme reagent, a substrate solution is added to the wells and color develops in proportion to the amount of XRCC2 bound in the initial step. The color development is stopped and the intensity of the color is measured.
Product Overview
XRCC2participates in homologous recombination to maintain chromosome stability and repair DNA damage. This gene is involved in the repair of DNA double-strand breaks by homologous recombination and it functionally complements Chinese hamster irs1, a repair-deficient mutant that exhibits hypersensitivity to a number of different DNA-damaging agents.Hamster cells deficient in XRCC2 showed a more than 100-fold decrease in homologous recombination induced by double-strand breaks compared with the parental cell line. This defect was corrected to almost wildtype levels by transient transfection with a plasmid expressing XRCC2. The repair defect in XRCC2 mutant cells appeared to be restricted to recombinational repair because nonhomologous end joining was normal.
Components
Reagents | Quantity | Reagents | Quantity |
Assay plate (96 Wells) | 1 | Instruction manual | 1 |
Standard (lyophilized) | 2 | Sample Diluent | 1 x 20 mL |
Biotin-Conjugate (concentrate 100 x) | 1 x 120 μL | Biotin-Conjugate Diluent | 1 x 12 mL |
Streptavidin-HRP (concentrate 100 x) | 1 x 120 μL | Streptavidin-HRP Diluent | 1 x 12 mL |
Wash Buffer (concentrate 25 x) | 1 x 20 mL | Substrate Solution | 1 x 10 mL |
Stop Solution | 1 x 6 mL | Adhesive Films | 4 |
Specificity
This assay has high sensitivity and excellent specificity for detection of Human XRCC2. No significant cross-reactivity or interference between Human XRCC2 and analogues was observed.
Recovery
Matrices listed below were spiked with certain level of recombinant Human XRCC2 and the recovery rates were calculated by comparing the measured value to the expected amount of Human XRCC2 in samples.
Sample Type | Number | Recovery range (%) | Average(%) |
Serum | 10 | 90-101 | 96 |
EDTA plasma | 10 | 89-97 | 93 |
Heparin plasma | 10 | 91-99 | 95 |
Precision
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in forty separate assays to assess inter-assay precision. CV (%) = SD/meanX100 Intra-Assay: CV<8% Inter-Assay: CV<12%
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Human XRCC2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample Type | 1:2 | 1:4 | 1:8 | 1:16 |
Serum | 78-89% | 81-99% | 92-103% | 95-105% |
EDTA plasma | 91-101% | 90-98% | 93-101% | 91-98% |
Heparin plasma | 92-103% | 93-102% | 92-99% | 91-101% |
Stability
The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition. The loss rate was determined by accelerated thermal degradation test. Keep the kit at 37°C for 4 and 7 days, and compare O.D.values of the kit kept at 37°C with that of at recommended temperature. (referring from China Biological Products Standard, which was calculated by the Arrhenius equation. For ELISA kit, 4 days storage at 37°C can be considered as 6 months at 2 - 8°C, which means 7 days at 37°C equaling 12 months at 2 - 8°C).
Sample collection and storage
Serum: Use a serum separator tube (SST) and allow samples to clot for two hours at room temperature or overnight at 2 - 8°C before centrifugation for 15 minutes at 1000 × g. Remove serum and assay immediately or aliquot and store samples at ≤ -20°C. Avoid repeated freeze-thaw cycles. Plasma: Collect plasma using EDTA, or heparin as an anticoagulant. Centrifuge for 15 minutes at 1000 × g at 2 - 8°C within 30 minutes of collection. Assay immediately or aliquot and store samples at ≤ -20°C. Avoid repeated freeze-thaw cycles. Other biological fluids: Centrifuge samples for 20 minutes at 1000 × g. Remove particulates and assay immediately or store samples in aliquot at -20°C or -80°C. Avoid repeated freeze/thaw cycles.
Kits storage instructions